S4-HCLThe hydrogen chloride gas detection sensor is a constant potential electrolytic sensor. The hydrogen chloride and oxygen will undergo corresponding redox reaction on the working electrode and the counter electrode respectively and release the charge to form a current. The generated current is proportional to the concentration of hydrogen chloride. The concentration of hydrogen chloride can be determined by measuring the current.
Item | Parameters |
model | S4-HCL |
Detection principle | Electrochemical Three electrodes |
Detection gas | HCL |
Detection range | 0-50PPM |
Maximum load concentration | 100PPM |
output signal | 300±100nA/PPM |
Zero drift | 0~5PPM |
Resolution | 0.1PPM |
Response time | <70s |
Bias voltage | +200mv |
Load resistance | 5~30Ω |
Working Temperature | -30℃-50℃ |
Working Humidity | 15%RH to 90%RH(no condensation) |
Repeatability | <±2% output signal |
Long-term stability | <2% signal/month |
Linearity | Linearity regression coefficient R²=0.999 |
Work pressure | 90 to 110 kPa |
Shelf life | 12 months after delivery |
Service life | 2 years |